CONFIGURING THE METER

USING SETPOINT 1 CONFIGURATION ("S1.CF")
Setpoint 1 Configuration (S1.CF) is not active unless your meter has dual relay output capabilities. The LED's will display whether the ("S1.CF") is active or not. You may use Setpoint 1 Configuration ("S1.CF") for the following:
    * Setting setpoint's active band above or below your chosen value
    * Selecting whether the setpoint operation is latched or unlatched
    * Assigning setpoint values to the net or gross reading
Setting Setpoint 1's Active Band
1. Press the MENU button until the meter shows "S1.CF".

2. Press the /TARE button. The meter shows one of the following:
* "S.1=A" (Active above the deadband) (default)
* "S.1=B" (Active below the deadband)

3. Press the /NT/GRS button to toggle between available choices.

4. Press the /TARE button to select if Setpoint 1 is latched or unlatched or press the MENU button to store your selection and enter "S2.CF" (Setpoint 2 Configuration)

Selecting if Setpoint 1 is Latched or Unlatched
1. Press the /TARE button. The meter shows one of the following:
* "S.2=L" (setpoint 1 to be latched) (default)
* "S.2=U" (setpoint 1 to be unlatched)

2. Press the /NT/GRS button to toggle between available choices.

3. Press the /TARE button to assign Setpoint 1 values to net or gross reading or press the MENU button to enter "S2.CF" (Setpoint 2 Configuration).

Assigning Setpoint 1 Values to Net or Gross Readings
1. Press the /TARE button. The meter shows one of the following:
* "S.3=N" setpoint 1 assigned to net reading (default)
* "S.3=G" setpoint 1 assigned to gross reading

2. Press the /NT/GRS button to toggle between available choices.

3. Press the MENU button to store your selection(s). The meter momentarily shows "STRD", followed by "S2.CF" (Setpoint 2 Configuration).

USING SETPOINT 2 CONFIGURATION ("S2.CF")
Setpoint 2 Configuration ("S2.CF") is not active unless your meter has dual relay output capabilities. The LED's will display whether the ("S2.CF") is active or not. You may use Setpoint 2 Configuration ("S2.CF") for the following:
    * To set setpoint's active band above or below your chosen value
    * To select whether the setpoint operation is latched or unlatched
    * To assign setpoint values to the net or gross reading
Setting Setpoint 2's Active Band
1. Press the MENU button until the meter shows "S2.CF".


2. Press the /TARE button. The meter shows one of the following:
* "S.1=A" (Active above the deadband) (default)
* "S.1=B" (Active below the deadband)

3. Press the /NT/GRS button to toggle between available choices.

4. Press the /TARE button to select if Setpoint 1 is latched or unlatched or press the MENU button to store your selection and enter "S1.DB" (Setpoint 1 Deadband)

Selecting if Setpoint 2 is Latched or Unlatched
1. Press the /TARE button. The meter shows one of the following:
* "S.2=L" (setpoint 1 to be latched) (default)
* "S.2=U" (setpoint 1 to be unlatched)

2. Press the /NT/GRS button to toggle between available choices.

3. Press the /TARE button to assign Setpoint 2 values to net or gross reading or press the MENU button to enter "S1.DB" (Setpoint 1 Deadband).

Assigning Setpoint 2 Values to Net or Gross Readings
1. Press the /TARE button. The meter shows one of the following:
* "S.3=N" (setpoint 2 assigned to net reading) (default)
* "S.3=G" (setpoint 2 assigned to gross reading)

2. Press the /NT/GRS button to toggle between available choices.

3. Press the MENU button to store your selection(s). The meter momentarily shows "STRD", followed by "S1.DB" (Setpoint 1 Deadband).

SETTING THE SETPOINT 1 DEADBAND ("S1.DB")
Setpoint 1 Deadband ("S1.DB") is not active unless your meter has dual relay output capabilities. The LED's will display whether the ("S1.DB") is active or not. The Setpoint 1 default deadband is 0003. To change the deadband (hysteresis) of setpoint 1, follow these steps:

1. Press the MENU button until the meter shows "S1.DB".

2. Press the /TARE button. The meter shows the last previously stored 4-digit number (0000 through 9999) with flashing 4th digit.

3. Press the /NT/GRS button to change the value of the flashing digit. If you continue to press the /NT/GRS button, the flashing digit's value continues to change.

4. Press the /TARE button to scroll horizontally to the next digit.

5. Press the MENU button to store your selection. The meter momentarily shows "STRD", followed by "S2.DB" (Setpoint 2 Deadband).


SETTING THE SETPOINT 2 DEADBAND ("S2.DB")
Setpoint 2 Deadband ("S2.DB") is not active unless your meter has dual relay output capabilities. The LED's will display whether the (S2.DB) is active or not. The Setpoint 2 default deadband is 0003. To change the deadband (hysteresis) of setpoint 2, follow these steps:

1. Press the MENU button until the meter shows "S2.DB".

2. Press the /TARE button. The meter shows the last previously stored 4-digit number (0000 through 9999) with flashing 4th digit.

3. Press the /NT/GRS button to change the value of the flashing digit. If you continue to press the /NT/GRS button, the flashing digit's value continues to change.

4. Press the /TARE button to scroll horizontally to the next digit.

5. Press the MENU button to store your selection. The meter momentarily shows "STRD", followed by "OT.CF" (Output Configuration).